| sonic
      data
      processing 1. Instrument
      description and operation modes 1. Instrument description and operation modes During
      the BUBBLE experiment a total of 23 ultrasonic thermometer-anemometers (sonics)
      were operated: 5 R2 (4 omnidirectional / 1 asymmetric), 12 METEK USA-1, 4
      CSI CSAT3, 1 Gill HS and 1 Young 81000. All sonics of the same type were
      set - whenever possible - to the same settings described below: 
 (1) Except
      instruments at  BMES and VLNF
      operated at 10Hz output (60 Hz internal) with a datalogger. >
      Detailed
      description of all sensors (PDF, without
      Young 81000). 
 
 At VLNF and BMES the sonic raw-data was processed on site by dataloggers (Campbell 21x at VLNF, Campbell 23x at BMES, using SDM). At these 2 sites only 10min averages and covariances were stored. If available, a 2d matrix correction was applied to the raw data, to minimize effects of flow distortion. Refer to the description of wind tunnel calibrations during BUBBLE for detailed information on the matrices. The following table lists all instruments where 2d matrices are applied: 
 (1) Matrices NOT applied yet (October 20, 2002) For Gill R2 Sonics without a 2d-matrix, the manufacturer and instrument-individual "Gill correction" was applied: 
 No additional calibration was applied to the instruments listed below: 
 4.1 Internal Raw Data Check CSI
      CSAT3
      serial data are internally flagged by the instrument for bad records (i.e.
      rain drops). In
      the post processing bad 20 Hz records are linearly
      interpolated. If more than 256 CSAT3 records  during 10 minutes are flagged (approx. 12 seconds, 
      2%), then the 10min-block is completely discarded and no further statistics were
      computed. 4.2 Missing Records The LabView data aquisition (2)
      checks for dropped records. If there are missing records in a signal, the
      integral statistics can still be calculated with the remaining
      measurements of the block (e.g. with 35900 instead of 36000 values for
      half an hour, the flux is not substantially altered). Because the number of missing records in sequence is unknown, the
      gaps cannot be interpolated. 4.3 Despiking A simple despiking test is run over the 20 Hz data of all sonics including analog inputs. The despiking should avoid effects of single values that are completely off (called "spikes", caused by rain,dust,insects,other errors). The despiking-test checks every single 20Hz value zi of all parameters (z=u,v,w,t,q,CO2) to be within a range: zi
      > mean (z) - s(z) * a 
      a was empirically set a = 6 for u and v, a = 10 for w, t,and q and a = 20 for CO2.
       5.1 Krypton hygrometers A
      total of 6 CSI KH2O hygrometers were combined with sonics to determine
      latent heat fluxes during BUBBLE. The KH20s were scaled using the
      following calibrations: 
 
 Two Licor 7500 CO2/H2 at BSPR to provide information on Carbon Dioxide Exchange and latent heat flux. Both instruments were sampled with a time offset of 250 msec relatively to the sonic. Due to a software error of the Licor 7500 this offset does not correspond to the real offset. > Read our technical report on the " Impact of the LICOR 7500 Lag Correction on Field Data sampled 2000-2003 at the University of Basel" In
      the postprocessing the offset is shifted back. The signal was transferred using the analog
      output of the LICOR 7500 and feeding it into the analog inputs of the
      sonics (Gill HS and Gill R2). The instrument at 14.7m was
      additionally sampled serially. 
 (A) Settings before 26/06/02, 08:35, (B) Settings after 26/06/02, 08:35. 6. Processing of Mean Statistics 6.1 Coordinate System The coordinate systems of all sonics are rotated so that u+ points towards geographic East and v+ points towards geographic N. w+ always points upward. Note that, this is no rotation into mean wind i.e. no streamline rotation was applied. 6.2 General Statistics Statistiscs
    were calculated from simple blocks over 10 minutes without detrending. Statistical moments are calculated for all parameters (u,v,w,t,q,c)
      as well as the covariance-matrix for u,v,w,t,q,c All
      output parameters are
      checked to be within a range. Values that are out of this range are
      removed and set to an error value. > List of
    range- and clipping-settings for all parameters. Flux densities labeled "corrected" in the
      database are: QH QE QS Q All
      turbulent fluxes (QH,QE,from block
      averages of 20 Hz raw data over one hour without detrending. All
      fluxes are vertically oriented and no run-to-run streamline rotation was
      applied.  7.1
      Sensible heat flux density Q QH was calculated from the 60 min block value of the covariance w'T' (vertical wind / acoustic temperature). Air density and heat capacity are calculated for each time step based upon measurements of air temperature, humidity and air pressure at the sites. The following correction were applied in the indicated order: 
 7.2 Latent heat flux density QE 
	QE
	was
      calculated from the 60 min block value of the covariance
      
 w'a' (vertical wind / absolute humidity) Latent heat of vaporization, air density
      and heat capacity are calculated for each time step based upon
      measurements of air temperature, humidity and air pressure at the sites.
      The following correction were applied in the indicated order:
      
       Oxygen-Correction with instrument individual factors
          (only for KH20 and not the LICOR 7500, after Tanner et al. ,1983).
          The correction typically enlarges QE 
          between +1% and +12 %. In relative numbers it is more pronounced at
          the urban sites (absolute lower QE) Spectral correction and correction for sensor
          separation (Moore, 1986). The influence on QE is
          between 2% and 7%. WPL-correction (Webb et al., 1980). The relative
          influence on QE is between 2% and 25%. 7.3
      Storage / soil heat flux density  QS GRNZ, VLNF,
      BLER, GEMP)
      was measured directly by the mean of three heat flux plates between 3 and
      5 cm depth. Storage in the vegetation was neglected. At BMES,
      two experimental heat flux plates were placed directly above the
      horizontal concrete surface. The following correction was applied:
       QS
          was corrected for flux density divergence in the soil
          level above the plates using measured soil temperatures. Soil density
          was set to 1300 kg/m3 and the soil heat capacity to a constant value
          of 1300 J kg^-1 K^-1. This correction was not applied to BMES. 7.4 Flux Corrections applied to the mass flux density
      of QCO2
      
       The CO2-Flux is calculated with the
          WPL-Correction. (Webb et al., 1980). Spectra and cospectra are calculated over 1h runs. 
      
       8.1 Data Checks
      
        The
      preprocessing and data checks include all features described above (see 3
      for matrix correction or Gill manufacturer correction and 4
      for data checks, missing records and  despiking). There are additional procedures and data
      checks carried out before the FFT is applied:
       The coordinate system is rotated into mean horizontal
          wind by a single rotation around the z-axis (i.e. makes mean(v)
          equal 0, but mean(w) must not be 0). In the street canyon and the
          near-roof levels, vertical winds (w) are physically not zero in mean,
          therefore the usually applied second rotation or a
          planar fit would make no sense. All data rows (u,v,w,t,q,c) are  linearly detrended
          before applying the FFT. Note that the detrending is only applied to the
          data for the FFT but not for the integral turbulence statistics described in 6.
          Detrending alters the energy by removing energy, therfore be careful
          when budgeting energy (TKE). Both, the instantaneous values and standard deviation of
          the whole block must be within a defined range. 8.2 FFT Spectra and Cospectra
    are calculated in a resolution of 64 spectral bands. The spectral bands are
    logarithmic equally spaced between 0.0003 Hz (55 min) and 10 Hz. The same
    band definition is applied to all sonics and all stations to allow a fast
    and easy averaging and intercomparison. The last 0-255
    records of a 1h block (up to 12 seconds) are discarded to get a record
    number that is a multiple of 256 in order to to accelerate the FFT. References
      
    
      
       Schotanus,
      P., Nieuwstadt, F., and de Bruin, H. (1983):"Temperature
      measurement with a sonic anemometer and its application to heat and
      moisture fluxes".
      
		Boundary-Layer
      Meteorol., 26:81-93. Tanner,
      B. D., Swiatek, E., Greene, J. P. (1993): “Density fluctuations and use
      of the Krypton Hygrometer in surface flux measurements”. 
		Management of irrigation and drainage systems,  
      Webb, E.,
      Pearman, G., Leuning, R. (1980):
      “Correction of flux measurements for density effects due to heat and
      water wapour transfer”, Quart. J.
      Roy. Meteorol. Soc., 106,
      pp 85-100. |  |   
 |  | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||